<?xml version="1.0" encoding="UTF-8"?><!DOCTYPE article PUBLIC "-//NLM//DTD JATS (Z39.96) Journal Publishing DTD v1.2 20190208//EN" "http://jats.nlm.nih.gov/publishing/1.2/JATS-journalpublishing1.dtd"><article xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink" article-type="research-article" dtd-version="1.2" xml:lang="en">
    <front>
        <journal-meta>
            <journal-id journal-id-type="pmc">F1000Research</journal-id>
            <journal-title-group>
                <journal-title>F1000Research</journal-title>
            </journal-title-group>
            <issn pub-type="epub">2046-1402</issn>
            <publisher>
                <publisher-name>F1000 Research Limited</publisher-name>
                <publisher-loc>London, UK</publisher-loc>
            </publisher>
        </journal-meta>
        <article-meta>
            <article-id pub-id-type="doi">10.12688/f1000research.110966.1</article-id>
            <article-categories>
                <subj-group subj-group-type="heading">
                    <subject>Research Article</subject>
                </subj-group>
                <subj-group>
                    <subject>Articles</subject>
                </subj-group>
            </article-categories>
            <title-group>
                <article-title>Establishment of diagnostic reference level and radiation dose variation in head &amp; neck and pelvis treatment planning in radiation therapy computed tomography</article-title>
                <fn-group content-type="pub-status">
                    <fn>
                        <p>[version 1; peer review: awaiting peer review]</p>
                    </fn>
                </fn-group>
            </title-group>
            <contrib-group>
                <contrib contrib-type="author" corresp="no">
                    <name>
                        <surname>Rao</surname>
                        <given-names>Ms. Shreekripa</given-names>
                    </name>
                    <role content-type="http://credit.niso.org/">Data Curation</role>
                    <role content-type="http://credit.niso.org/">Investigation</role>
                    <role content-type="http://credit.niso.org/">Project Administration</role>
                    <role content-type="http://credit.niso.org/">Writing &#x2013; Original Draft Preparation</role>
                    <xref ref-type="aff" rid="a1">1</xref>
                </contrib>
                <contrib contrib-type="author" corresp="yes">
                    <name>
                        <surname>Kadavigere</surname>
                        <given-names>Dr. Rajagopal</given-names>
                    </name>
                    <role content-type="http://credit.niso.org/">Conceptualization</role>
                    <role content-type="http://credit.niso.org/">Investigation</role>
                    <role content-type="http://credit.niso.org/">Methodology</role>
                    <role content-type="http://credit.niso.org/">Project Administration</role>
                    <role content-type="http://credit.niso.org/">Supervision</role>
                    <uri content-type="orcid">https://orcid.org/0000-0003-3486-8740</uri>
                    <xref ref-type="corresp" rid="c1">a</xref>
                    <xref ref-type="aff" rid="a2">2</xref>
                </contrib>
                <contrib contrib-type="author" corresp="no">
                    <name>
                        <surname>Sharan</surname>
                        <given-names>Dr. Krishna</given-names>
                    </name>
                    <role content-type="http://credit.niso.org/">Formal Analysis</role>
                    <role content-type="http://credit.niso.org/">Funding Acquisition</role>
                    <role content-type="http://credit.niso.org/">Methodology</role>
                    <role content-type="http://credit.niso.org/">Resources</role>
                    <xref ref-type="aff" rid="a3">3</xref>
                </contrib>
                <contrib contrib-type="author" corresp="no">
                    <name>
                        <surname>Sukumar</surname>
                        <given-names>Suresh</given-names>
                    </name>
                    <role content-type="http://credit.niso.org/">Conceptualization</role>
                    <role content-type="http://credit.niso.org/">Data Curation</role>
                    <role content-type="http://credit.niso.org/">Methodology</role>
                    <uri content-type="orcid">https://orcid.org/0000-0001-9345-9790</uri>
                    <xref ref-type="aff" rid="a4">4</xref>
                </contrib>
                <contrib contrib-type="author" corresp="no">
                    <name>
                        <surname>GC</surname>
                        <given-names>Mr. Srinidhi</given-names>
                    </name>
                    <role content-type="http://credit.niso.org/">Data Curation</role>
                    <role content-type="http://credit.niso.org/">Writing &#x2013; Original Draft Preparation</role>
                    <uri content-type="orcid">https://orcid.org/0000-0001-9717-5410</uri>
                    <xref ref-type="aff" rid="a3">3</xref>
                </contrib>
                <contrib contrib-type="author" corresp="no">
                    <name>
                        <surname>Dsouza</surname>
                        <given-names>Mrs. Rechal Nisha</given-names>
                    </name>
                    <role content-type="http://credit.niso.org/">Formal Analysis</role>
                    <role content-type="http://credit.niso.org/">Writing &#x2013; Review &amp; Editing</role>
                    <xref ref-type="aff" rid="a1">1</xref>
                </contrib>
                <contrib contrib-type="author" corresp="no">
                    <name>
                        <surname>S</surname>
                        <given-names>Mr. Suman</given-names>
                    </name>
                    <role content-type="http://credit.niso.org/">Data Curation</role>
                    <role content-type="http://credit.niso.org/">Writing &#x2013; Review &amp; Editing</role>
                    <xref ref-type="aff" rid="a1">1</xref>
                </contrib>
                <aff id="a1">
                    <label>1</label>Department of Radiotherapy and Oncology, Manipal College of Health Professions, Manipal, Karnataka, 576104, India</aff>
                <aff id="a2">
                    <label>2</label>Department of Radiodiagnosis and Imaging, Kasturba Medical College and Hospital, Manipal, Karnataka, 576104, India</aff>
                <aff id="a3">
                    <label>3</label>Department of Radiotherapy and Oncology, Kasturba Medical College and Hospital, Manipal, Karnataka, 576104, India</aff>
                <aff id="a4">
                    <label>4</label>Department of Medical Imaging Technology, Manipal College of Health Professions, Manipal, Karnataka, 576104, India</aff>
            </contrib-group>
            <author-notes>
                <corresp id="c1">
                    <label>a</label>
                    <email xlink:href="mailto:rajarad@gmail.com">rajarad@gmail.com</email>
                </corresp>
                <fn fn-type="conflict">
                    <p>No competing interests were disclosed.</p>
                </fn>
            </author-notes>
            <pub-date pub-type="epub">
                <day>3</day>
                <month>5</month>
                <year>2022</year>
            </pub-date>
            <pub-date pub-type="collection">
                <year>2022</year>
            </pub-date>
            <volume>11</volume>
            <elocation-id>489</elocation-id>
            <history>
                <date date-type="accepted">
                    <day>27</day>
                    <month>4</month>
                    <year>2022</year>
                </date>
            </history>
            <permissions>
                <copyright-statement>Copyright: &#x00a9; 2022 Rao MS et al.</copyright-statement>
                <copyright-year>2022</copyright-year>
                <license xlink:href="https://creativecommons.org/licenses/by/4.0/">
                    <license-p>This is an open access article distributed under the terms of the Creative Commons Attribution Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.</license-p>
                </license>
            </permissions>
            <self-uri content-type="pdf" xlink:href="https://f1000research.com/articles/11-489/pdf"/>
            <abstract>
                <p>
                    <bold>Background</bold>: This observational study aims to establish a discrepancy in the radiation dose distributed for head &amp; neck and pelvis localization in Computed Tomography (CT) imaging in radiation therapy. The objective and need of the current study are to establish Diagnostic Reference Level (DRL) for head &amp; neck and pelvic CT protocols in Radiation Therapy planning and compare them with other regional, national and international DRL. The established DRL will be used to find further optimized DRL for Radiation Therapy Computed Tomography (RTCT).</p>
                <p>
                    <bold>Methods</bold>: A total of 120 patients with Head &amp; Neck cancers and 90 patients with pelvic cancers to be treated with radiotherapy prescribed for the RTCT with ages above 18 years old and above were included in the current study. The advanced Philips 16 slice big bore CT acquired all the CT simulation images.</p>
                <p>
                    <bold>Results:</bold> Third quartile standards of Dose Length Product and effective dose for Head &amp; Neck, and pelvis were 790.65mGy.cm, 2.45179 mSv, and 999.7 mGy.cm, 15.48mSv respectively. The third quartile CTDI
                    <sub>vol</sub> value for the same order of procedure is 17.6 mGy. Tube voltage of 120kVp and 300mAs and used for radiotherapy planning.</p>
                <p>
                    <bold>Conclusion:</bold> The first regional Radiation therapy Computed tomography simulation Diagnostic reference levels have been projected and deliver a platform for dose evaluation and optimization due to the limited number of papers on radiation therapy, computed tomography, and diagnostic reference levels. Comparison with previously available RT CT diagnostic reference levels indicated some radiation dose variation, so exposure parameters should be revised and improved.</p>
            </abstract>
            <kwd-group kwd-group-type="author">
                <kwd>Radiotherapy</kwd>
                <kwd>Diagnostic reference levels</kwd>
                <kwd>Head &amp; Neck</kwd>
                <kwd>Pelvis</kwd>
                <kwd>DLP</kwd>
                <kwd>and CTDIvol</kwd>
            </kwd-group>
            <funding-group>
                <award-group id="fund-1">
                    <funding-source>Indian Council of Medical Research</funding-source>
                    <award-id>5/13/48/2020/NCD-III</award-id>
                </award-group>
                <funding-statement>The author(s) declared that no grants were involved in supporting this work.</funding-statement>
            </funding-group>
        </article-meta>
    </front>
    <body>
        <sec id="sec1" sec-type="intro">
            <title>Introduction</title>
            <p>Computed tomography (CT) plays a significant part in the radiotherapy treatment process. CT enables personalized radiation treatment delivery with the help of cross-sectional imaging. CT is the only modality used to calculate 3D-dose in radiotherapy.
                <sup>
                    <xref ref-type="bibr" rid="ref1">1</xref>
                </sup> Additionally, it facilitates inaccurate delivery of radiotherapy by obtaining digitally reconstructed radiographs for setup verification of the patient on the teletherapy machine before starting treatment.</p>
            <p>Despite these advantages, CT carries a significant disadvantage of utilizing ionizing radiation for imaging, with the inevitable risk of stochastic effects. As mentioned in the ALARA principle, these imaging doses, which patient organs will receive during the treatment, should be as low as possible. The &#x201c;International Commission on Radiological Protection&#x201d; (ICRP) implemented the &#x201c;Diagnostic Reference Level&#x201d; (DRL) used in radiological procedures. A Diagnostic reference level is not a regulatory limit. It is a benchmark that, when exceeded, triggers a review. ICRP recommends that the DRL represent the local practice within a particular geographical area. DRL will not distinguish between good or bad medical practices. The motive of establishing a DRL is to reduce the dose of ionizing radiation that does not contribute to giving any additional clinical information to a medical practitioner. DRL is estimated to improve the radiation dose and image quality.
                <sup>
                    <xref ref-type="bibr" rid="ref2">2</xref>
                </sup>
            </p>
            <p>DRLs are not applicable in radiotherapy practice, but it is necessary to apply DRL for different imaging techniques used in radiotherapy treatment, imaging for patient position verification, and treatment planning. DRL is the 75
                <sup>th</sup> percentile of effective dose distribution, calculated with patient or phantom data. Regional, national and international levels DRLs may be estimated and compared.
                <sup>
                    <xref ref-type="bibr" rid="ref2">2</xref>
                </sup> We need to periodically estimate DRLs as there may be a change in the work practice.</p>
            <p>Diagnostic reference level for CT procedure is already available, and studies will still establish diagnostic reference levels regionally and nationally. For radiotherapy planning CT, the scanning parameters and the scan length for each region are different. It is essential to establish a DRL for the same. Some studies on establishing DRL for radiotherapy planning CT have already been published.
                <sup>
                    <xref ref-type="bibr" rid="ref1">1</xref>
                </sup>
                <sup>&#x2013;</sup>
                <sup>
                    <xref ref-type="bibr" rid="ref3">3</xref>
                </sup> The objective and need of this study are to establish diagnostic reference levels for head &amp; neck and pelvic CT protocols in RT treatment planning and compare them with other regional, national and international DRL. The established DRL will be used to find further optimized DRL for RT CT.</p>
            <p>An image of acceptable quality can be obtained using a minimum radiation dose. The present study was conducted using a phantom to assess the radiographic technical parameters that could lead to the lowest radiation dose and acceptable image quality. The outcomes of this study can be used to improve Head &amp; neck, and pelvis radiographic examinations.</p>
        </sec>
        <sec id="sec2" sec-type="methods">
            <title>2. Methods</title>
            <p>This study was conducted after obtaining approval from the ethical committee of Kasturba (IEC number 925/2018) Hospital. The expected outcome of the current study is &#x201c;standardization of Radiotherapy Planning CT Head and Neck and Pelvic protocol and optimization of dose can be achieved by using reference DRLs.&#x201d; By using a convenient sampling technique and Kappa co-efficient, a total of 120 patients with head and neck cancer and 90 patients with pelvic cancer to be treated with radiotherapy 
                <strike>(</strike>Mean age 53.33 &#x00b1;8.5, Male 54.16 &#x00b1; 15 Female 52.5&#x00b1;12.17) were prescribed for RTCT with the age of 18 years and above were taken for the study after obtaining a voluntary written informed consent form from February 2019-August 2019 All the CT simulation images were acquired in the advanced Philips 16 slice big-bore RT CT machine, with a tube voltage of 120 kVp and 300 mAs and used for radiotherapy1 planning. During the prospective study, information such as the patient&#x2019;s gender, region of interest, Field-Of-View (FOV), type of cancer, DLP, and CTDI
                <sub>vol</sub> were noted. The examination protocol and the exposure parameters used to take the scan were also collected. The data collected for this study was anonymized entirely.</p>
            <p>In this study, we have included and analysed CT procedures of the Head &amp; Neck, and Pelvis. The CT image acquisition parameters are given in 
                <xref ref-type="table" rid="T1">Table 1</xref>.</p>
            <table-wrap id="T1" orientation="portrait" position="float">
                <label>Table 1. </label>
                <caption>
                    <title>The examination parameters.</title>
                </caption>
                <table content-type="article-table" frame="hsides">
                    <thead>
                        <tr>
                            <th align="left" colspan="1" rowspan="1" valign="top">Location</th>
                            <th align="left" colspan="1" rowspan="1" valign="top">Acquisition type</th>
                            <th align="left" colspan="1" rowspan="1" valign="top">Voltage (kV)</th>
                            <th align="left" colspan="1" rowspan="1" valign="top">Reference (mAs)</th>
                            <th align="left" colspan="1" rowspan="1" valign="top">Collimation (N&#x00d7;mm)</th>
                            <th align="left" colspan="1" rowspan="1" valign="top">Rotation Time (sec)</th>
                            <th align="left" colspan="1" rowspan="1" valign="top">Pitch Slice</th>
                            <th align="left" colspan="1" rowspan="1" valign="top">Thickness (mm)</th>
                        </tr>
                    </thead>
                    <tbody>
                        <tr>
                            <td align="left" colspan="1" rowspan="1" valign="top">
                                <bold>RT Head &amp; Neck</bold>
                            </td>
                            <td align="left" colspan="1" rowspan="1" valign="top">Helical</td>
                            <td align="left" colspan="1" rowspan="1" valign="top">120</td>
                            <td align="left" colspan="1" rowspan="1" valign="top">300</td>
                            <td align="left" colspan="1" rowspan="1" valign="top">16 &#x00d7; 1.5</td>
                            <td align="left" colspan="1" rowspan="1" valign="top">1</td>
                            <td align="left" colspan="1" rowspan="1" valign="top">0.813</td>
                            <td align="left" colspan="1" rowspan="1" valign="top">3</td>
                        </tr>
                        <tr>
                            <td align="left" colspan="1" rowspan="1" valign="top">
                                <bold>RT Pelvis</bold>
                            </td>
                            <td align="left" colspan="1" rowspan="1" valign="top">Helical</td>
                            <td align="left" colspan="1" rowspan="1" valign="top">120</td>
                            <td align="left" colspan="1" rowspan="1" valign="top">300</td>
                            <td align="left" colspan="1" rowspan="1" valign="top">16 &#x00d7; 1.5</td>
                            <td align="left" colspan="1" rowspan="1" valign="top">1</td>
                            <td align="left" colspan="1" rowspan="1" valign="top">0.813</td>
                            <td align="left" colspan="1" rowspan="1" valign="top">5</td>
                        </tr>
                    </tbody>
                </table>
                <table-wrap-foot>
                    <p>kVp - kilovoltage peak, mAs - milli-ampere-second.</p>
                </table-wrap-foot>
            </table-wrap>
            <p>Scanning length per examination varied from one scan to another scan. The considered length for head &amp; neck investigation was from the vertex of the Head to the carina, and for the pelvis, the procedure is from the D10-D12 area to the mid femur. The &#x201c;American Association of Physicists in Medicine suggests that the scanned volume should extend at least 5 cm superiorly and inferiorly beyond the target area&#x201d;.
                <sup>
                    <xref ref-type="bibr" rid="ref4">4</xref>
                </sup>
            </p>
            <p>All RTCT scan images were verified and approved by the oncologist after confirming them to be suitable for volume delineation. The effective dose was arrived at using the Normalized effective dose (k) coefficients and K&#x2019;s value according to the updated AAPM report 96.</p>
            <p>Patient weight 40 &#x2013; 80 kg mean weight 60 kg &#x00b1; 5 kg</p>
            <sec id="sec3">
                <title>2.1 Objective image quality assessment</title>
                <p>The obtained CT image of a different patient was analyzed for dose and quantitate image quality. The &#x201c;signal-to-noise&#x201d; and &#x201c;contrast-to-noise&#x201d; ratios for other density regions were analyzed by sketching circular regions of interest (ROIs) of 20-22 mm
                    <sup>2</sup> and were manually positioned in the different areas and background to encompass the homogeneity of measured tissues.
                    <sup>
                        <xref ref-type="bibr" rid="ref3">3</xref>
                    </sup>
                </p>
                <p>The location of ROIs for quantitative image quality assessment of the pelvis and Head and neck is shown in 
                    <xref ref-type="fig" rid="f1">Figure 1A</xref> and 
                    <xref ref-type="fig" rid="f1">Figure 1B</xref>, respectively.</p>
                <fig fig-type="figure" id="f1" orientation="portrait" position="float">
                    <label>Figure 1. </label>
                    <caption>
                        <title>A: Location of Region of Interest (ROIs) for quantitative image quality assessment for the pelvis. B: Location of Region of Interest (ROIs) for quantitative image quality assessment for Head and neck.</title>
                    </caption>
                    <graphic id="gr1" orientation="portrait" position="float" xlink:href="https://f1000research-files.f1000.com/manuscripts/122632/23de5b71-49e4-4ac9-bb3c-2d58663a9937_figure1.gif"/>
                </fig>
                <p>Analysis of the SNR, CNR, and FOM ratios for the phantom and the participant population were arrived at by the following equation
                    <sup>
                        <xref ref-type="bibr" rid="ref5">5</xref>
                    </sup>:
                    <disp-formula id="e1">
                        <mml:math display="block">
                            <mml:mi>SNR</mml:mi>
                            <mml:mo>=</mml:mo>
                            <mml:mfenced close=")" open="(">
                                <mml:msub>
                                    <mml:mi>CT</mml:mi>
                                    <mml:mi>ROI</mml:mi>
                                </mml:msub>
                            </mml:mfenced>
                            <mml:mo>/</mml:mo>
                            <mml:msub>
                                <mml:mi>SD</mml:mi>
                                <mml:mi>ROI</mml:mi>
                            </mml:msub>
                            <mml:mspace width="0.25em"/>
                        </mml:math>
                    </disp-formula>
                    <disp-formula id="e2">
                        <mml:math display="block">
                            <mml:mi>CNR</mml:mi>
                            <mml:mo>=</mml:mo>
                            <mml:mfenced close=")" open="(">
                                <mml:mrow>
                                    <mml:msub>
                                        <mml:mi>CT</mml:mi>
                                        <mml:mi>ROI</mml:mi>
                                    </mml:msub>
                                    <mml:mo>&#x2013;</mml:mo>
                                    <mml:msub>
                                        <mml:mi>CT</mml:mi>
                                        <mml:mtext>Background</mml:mtext>
                                    </mml:msub>
                                </mml:mrow>
                            </mml:mfenced>
                            <mml:mo>/</mml:mo>
                            <mml:msub>
                                <mml:mi>SD</mml:mi>
                                <mml:mtext>Background</mml:mtext>
                            </mml:msub>
                            <mml:mspace width="0.25em"/>
                        </mml:math>
                    </disp-formula>
                    <disp-formula id="e3">
                        <mml:math display="block">
                            <mml:mi>FOM</mml:mi>
                            <mml:mo>=</mml:mo>
                            <mml:msup>
                                <mml:mi>CNR</mml:mi>
                                <mml:mn>2</mml:mn>
                            </mml:msup>
                            <mml:mo>/</mml:mo>
                            <mml:msub>
                                <mml:mtext>CTDI</mml:mtext>
                                <mml:mi>vol</mml:mi>
                            </mml:msub>
                        </mml:math>
                    </disp-formula>
                </p>
                <p>The statistics were analyzed using IBM SPSS STATISTICS 26.0 software. The descriptive statistical analysis included the average, standard deviation, median, and the third quartile for CTDI 
                    <sub>vol</sub> and total dose length product (DLP), calculated for each procedure.</p>
            </sec>
            <sec id="sec4">
                <title>2.2 Phantom study</title>
                <p>The maximum crucial part of dose optimization is completed earlier than the patient rests on the Computed tomography table, in the phrases of ensuring the &#x201c;justification of the clinical indication for CT.&#x201d; Numerous investigators have discovered different techniques for reducing radiation doses.
                    <sup>
                        <xref ref-type="bibr" rid="ref6">6</xref>
                    </sup>
                    <sup>&#x2013;</sup>
                    <sup>
                        <xref ref-type="bibr" rid="ref10">10</xref>
                    </sup>
                </p>
                <p>Once it is determined that a CT scan is required, the work of reducing radiation dose should confirm that pictures with interpretable indicative data can be achieved with the most reduced feasible radiation dose. A low dose CT scan without the required diagnostic information helps but could lead to further imaging or damage of significant time for producing management choices. &#x201c;In contrast, a high-dose CT with image quality superior to it needed to get diagnostic information could increase considerations over radiation dose risks.&#x201d;</p>
                <p>To observe a harmony between &#x201c;image quality and imaging dose, a CatPhan 500 phantom,&#x201d; which has a 20 cm width, was filtered. &#x201c;Scans were procured with a rotation time of 0.5 s, slice thickness of 3 mm, the pitch of 0.6, and the greatest accessible, effective tube current&#x201d;. With the use of reconstruction algorithms, the raw data were reconstructed from the phantom data. &#x201c;The reconstructed images of the CATPHAN from all diverse tube voltage and effective tube current-time product settings were reviewed by five experienced 
                    <ext-link ext-link-type="uri" xlink:href="https://www.sciencedirect.com/topics/medicine-and-dentistry/medical-physicist">medical physicists</ext-link>.&#x201d; &#x201c;To calculate the contrast-to-noise ratio, circular regions of interest were drawn with the signal ROI inside the 20 mm diameter target in the different density material and the background ROI outside but next to the target. To determine the appropriate tube voltage, the CNR, the increased dose, and penetration were considered&#x201d;.
                    <sup>
                        <xref ref-type="bibr" rid="ref11">11</xref>
                    </sup> Some of these parameters can optimize the dose and image quality.</p>
            </sec>
            <sec id="sec5">
                <title>2.3 Tube current</title>
                <p>Because of the rectilinear relationship between radiation dose &amp; applied tube current, tube current (measured in milliamperes) change is the utmost commonly used scan limit to modify dose.
                    <sup>
                        <xref ref-type="bibr" rid="ref12">12</xref>
                    </sup> Tube current (milliamperes) can be modified manually by selecting fixed or constant milliamperes with a user-required image quality measured to alter tube current based on body shape and body regions. While unnecessarily vigorous tube current reduction causes increased image noise in the mediastinum and chest wall, previous research has shown that tube current can be decreased to 15 to 50 mAs without compromising the diagnosis of lung or mediastinal anomalies.
                    <sup>
                        <xref ref-type="bibr" rid="ref13">13</xref>
                    </sup>
                </p>
            </sec>
            <sec id="sec6">
                <title>2.4 Tube potential</title>
                <p>Another scan parameter typically tailored for radiation exposure adaptation is tube potential (measured in kilovolts). The variation in radiation dose corresponds to tube current and is generally relative to the square of the adjustment of applied pinnacle kV. (kVp), for example, bringing down the tube potential from 140 to 120 kVp brings about a 35% reduction.</p>
            </sec>
            <sec id="sec7">
                <title>2.5 Scan length</title>
                <p>The total radiation dose associated with CT examination is the dose length product (DLP; measured in milli-gray multiplied by centimetres). For helical CT acquisitions, it is calculated as the product of volume CT dose index (CTDI
                    <sub>vol</sub>, in milli-gray) and scan length (in centimeters). Therefore, a reduction in scan length results in a direct and linear decrease in the DLP. Scan length should always be curtailed to the area of interest.
                    <sup>
                        <xref ref-type="bibr" rid="ref13">13</xref>
                    </sup>
                </p>
                <p>This experimental study was conducted in the same setup. The image acquisition was accomplished using a CATPHAN 502 phantom. These scans were used to relatively compare the quality of the obtained images using the technical parameters. Image acquisitions were obtained for the different mAs and kV settings to reduce radiation dose. As with tube current, lowering tube potential also improves image noise, improving image contrast. In some studies, it is mentioned that in common, chest CT can be performed at 80 kV in subjects smaller than 50 to 60 kg and 100 kV for subjects weighing up to 75 to 80 kg.
                    <sup>
                        <xref ref-type="bibr" rid="ref13">13</xref>
                    </sup>
                </p>
                <p>The improved image noise with decreased kilovolts does not influence general image quality considerably. Earlier investigations described weight-based streamlining of kilovolts and milliamperes for paediatric chest CT with a considerable decrease in dose.
                    <sup>
                        <xref ref-type="bibr" rid="ref13">13</xref>
                    </sup>
                </p>
                <p>A total of 100 images were acquired using the different technical parameters mentioned in 
                    <xref ref-type="table" rid="T2">Table 2</xref>. The acquired images were assessed for objective and subjective image quality acceptability by determining the SNR, CNR, and FOM.</p>
                <table-wrap id="T2" orientation="portrait" position="float">
                    <label>Table 2. </label>
                    <caption>
                        <title>Imaging parameters used for Phantom scan.</title>
                    </caption>
                    <table content-type="article-table" frame="hsides">
                        <thead>
                            <tr>
                                <th align="left" colspan="1" rowspan="1" valign="top">Imaging parameters</th>
                                <th align="left" colspan="1" rowspan="1" valign="top">Details</th>
                            </tr>
                        </thead>
                        <tbody>
                            <tr>
                                <td align="left" colspan="1" rowspan="1" valign="top">
                                    <bold>kVp</bold>
                                </td>
                                <td align="left" colspan="1" rowspan="1" valign="top">90, 129, 140</td>
                            </tr>
                            <tr>
                                <td align="left" colspan="1" rowspan="1" valign="top">
                                    <bold>mAs</bold>
                                </td>
                                <td align="left" colspan="1" rowspan="1" valign="top">Started with 100 till 350 with 50 mAs increment</td>
                            </tr>
                            <tr>
                                <td align="left" colspan="1" rowspan="1" valign="top">
                                    <bold>FOV</bold>
                                </td>
                                <td align="left" colspan="1" rowspan="1" valign="top">220 mm</td>
                            </tr>
                            <tr>
                                <td align="left" colspan="1" rowspan="1" valign="top">
                                    <bold>Scan length</bold>
                                </td>
                                <td align="left" colspan="1" rowspan="1" valign="top">183 mm</td>
                            </tr>
                            <tr>
                                <td align="left" colspan="1" rowspan="1" valign="top">
                                    <bold>Slice thickness</bold>
                                </td>
                                <td align="left" colspan="1" rowspan="1" valign="top">3 mm</td>
                            </tr>
                        </tbody>
                    </table>
                    <table-wrap-foot>
                        <p>kVp - kilovoltage peak, mAs - milli-ampere-second, FOV - Field of View.</p>
                    </table-wrap-foot>
                </table-wrap>
                <p>Scanning parameters of the phantom study with different kVp and mAs are shown in 
                    <xref ref-type="fig" rid="f2">Figures 2A</xref>, 
                    <xref ref-type="fig" rid="f2">2B</xref>, 
                    <xref ref-type="fig" rid="f2">2C</xref>, 
                    <xref ref-type="fig" rid="f2">2D</xref>, 
                    <xref ref-type="fig" rid="f3">3A</xref>, 
                    <xref ref-type="fig" rid="f3">3B</xref>, 
                    <xref ref-type="fig" rid="f3">3C</xref>, and 
                    <xref ref-type="fig" rid="f3">3D</xref>.</p>
                <fig fig-type="figure" id="f2" orientation="portrait" position="float">
                    <label>Figure 2. </label>
                    <caption>
                        <title>A: Phantom study with different 90 kVp and 300 mAs. B: Phantom study with different 120 kVp and 150 mAs. C: Phantom study with different 90 kVp and 250 mAs. D: Phantom study with different 90 kVp and 350 mAs.</title>
                    </caption>
                    <graphic id="gr2" orientation="portrait" position="float" xlink:href="https://f1000research-files.f1000.com/manuscripts/122632/23de5b71-49e4-4ac9-bb3c-2d58663a9937_figure2.gif"/>
                </fig>
                <fig fig-type="figure" id="f3" orientation="portrait" position="float">
                    <label>Figure 3. </label>
                    <caption>
                        <title>A: Phantom study with different 90 kVp and 250 mAs. B: Phantom study with different 140 kVp and 100 mAs. C: Phantom study with different 120 kVp and 300 mAs. D: Phantom study with different 140 kVp and 150 mAs.</title>
                    </caption>
                    <graphic id="gr3" orientation="portrait" position="float" xlink:href="https://f1000research-files.f1000.com/manuscripts/122632/23de5b71-49e4-4ac9-bb3c-2d58663a9937_figure3.gif"/>
                </fig>
                <p>Imaging parameters used in the phantom study are given in 
                    <xref ref-type="table" rid="T2">Table 2</xref>.</p>
            </sec>
        </sec>
        <sec id="sec8" sec-type="results">
            <title>3. Results</title>
            <p>In this study, we have used a total of 70 males and 50 female head and neck and 90 female patients that are diagnosed with cancer. The mean, median, effective dose, and third quartile values for DLP, which is the present protocol DRL for the Head and neck and pelvic scans, are summarised in 
                <xref ref-type="table" rid="T3">Table 3</xref>. All the subjects underwent the scan, and all the participants were eligible for potential reasons. All the 120 participants underwent scans without no missing data.</p>
            <table-wrap id="T3" orientation="portrait" position="float">
                <label>Table 3. </label>
                <caption>
                    <title>Table showing dose matrix &#x2013; Computed Tomography Dose Index (CTDI
                        <sub>vol</sub>) and Dose Length Product (DLP).</title>
                </caption>
                <table content-type="article-table" frame="hsides">
                    <thead>
                        <tr>
                            <th align="left" colspan="1" rowspan="1" valign="top">Region</th>
                            <th align="left" colspan="1" rowspan="1" valign="top"/>
                            <th align="left" colspan="1" rowspan="1" valign="top">Mean</th>
                            <th align="left" colspan="1" rowspan="1" valign="top">Median</th>
                            <th align="left" colspan="1" rowspan="1" valign="top">St. dev.</th>
                            <th align="left" colspan="1" rowspan="1" valign="top">3rd quartile (RT CT DRLs)</th>
                            <th align="left" colspan="1" rowspan="1" valign="top">CTDI
                                <sub>vol</sub> (mGy)</th>
                            <th align="left" colspan="1" rowspan="1" valign="top">Scan length (mm)</th>
                        </tr>
                    </thead>
                    <tbody>
                        <tr>
                            <td align="left" colspan="1" rowspan="2" valign="top">
                                <bold>H &amp; N</bold>
                            </td>
                            <td align="left" colspan="1" rowspan="1" valign="top">DLP (mGy cm)</td>
                            <td align="left" colspan="1" rowspan="1" valign="top">762.15</td>
                            <td align="left" colspan="1" rowspan="1" valign="top">760.35</td>
                            <td align="left" colspan="1" rowspan="1" valign="top">46.66</td>
                            <td align="left" colspan="1" rowspan="1" valign="top">790.65</td>
                            <td align="left" colspan="1" rowspan="4" valign="top">17.76</td>
                            <td align="left" colspan="1" rowspan="2" valign="top">383.88 &#x00b1; 28.77 (312 &#x2013; 453)</td>
                        </tr>
                        <tr>
                            <td align="left" colspan="1" rowspan="1" valign="top">Effective dose (mSv)</td>
                            <td align="left" colspan="1" rowspan="1" valign="top">2.36</td>
                            <td align="left" colspan="1" rowspan="1" valign="top">2.357</td>
                            <td align="left" colspan="1" rowspan="1" valign="top">0.14</td>
                            <td align="left" colspan="1" rowspan="1" valign="top">2.45179</td>
                        </tr>
                        <tr>
                            <td align="left" colspan="1" rowspan="2" valign="top">
                                <bold>Pelvis</bold>
                            </td>
                            <td align="left" colspan="1" rowspan="1" valign="top">DLP (mGy cm)</td>
                            <td align="left" colspan="1" rowspan="1" valign="top">929.24</td>
                            <td align="left" colspan="1" rowspan="1" valign="top">962.5</td>
                            <td align="left" colspan="1" rowspan="1" valign="top">96.97</td>
                            <td align="left" colspan="1" rowspan="1" valign="top">999.7</td>
                            <td align="left" colspan="1" rowspan="2" valign="top">492.5 &#x00b1; 35.5 (420 &#x2013; 580)</td>
                        </tr>
                        <tr>
                            <td align="left" colspan="1" rowspan="1" valign="top">Effective dose (mSv)</td>
                            <td align="left" colspan="1" rowspan="1" valign="top">13.93</td>
                            <td align="left" colspan="1" rowspan="1" valign="top">14.43</td>
                            <td align="left" colspan="1" rowspan="1" valign="top">1.454</td>
                            <td align="left" colspan="1" rowspan="1" valign="top">15.48</td>
                        </tr>
                    </tbody>
                </table>
                <table-wrap-foot>
                    <p>DLP - Dose Length Product, DRL - Diagnostic reference level, CTDI
                        <sub>vol</sub> - Volumetric computed tomography dose index.</p>
                </table-wrap-foot>
            </table-wrap>
            <p>The mean and standard deviation for image quality (SNR and CNR) and Figure of merit (FOM) for the Head &amp; Neck, and Pelvic scans are summarized in 
                <xref ref-type="table" rid="T4">Table 4</xref> and 
                <xref ref-type="table" rid="T5">Table 5</xref>.</p>
            <table-wrap id="T4" orientation="portrait" position="float">
                <label>Table 4. </label>
                <caption>
                    <title>Quantitative image parameter &#x2013; Head and Neck scans.</title>
                </caption>
                <table content-type="article-table" frame="hsides">
                    <thead>
                        <tr>
                            <th align="left" colspan="1" rowspan="1" valign="top">Region</th>
                            <th align="left" colspan="1" rowspan="1" valign="top">SNR Mean &#x00b1; SD (Range)</th>
                            <th align="left" colspan="1" rowspan="1" valign="top">CNR Mean &#x00b1; SD (Range)</th>
                            <th align="left" colspan="1" rowspan="1" valign="top">FOM Mean &#x00b1; SD (Range)</th>
                        </tr>
                    </thead>
                    <tbody>
                        <tr>
                            <td align="left" colspan="1" rowspan="1" valign="top">
                                <bold>CSF</bold>
                            </td>
                            <td align="left" colspan="1" rowspan="1" valign="top">2.28&#x00b1; 0.75 (0.94 &#x2013; 5.01)</td>
                            <td align="left" colspan="1" rowspan="1" valign="top">506.54&#x00b1;80.73 (282.35 &#x2013; 652.0854)</td>
                            <td align="left" colspan="1" rowspan="1" valign="top">14998.69 &#x00b1; 4176.01 (4519.84 &#x2013; 23968.69)</td>
                        </tr>
                        <tr>
                            <td align="left" colspan="1" rowspan="1" valign="top">
                                <bold>Caudate Nucleus</bold>
                            </td>
                            <td align="left" colspan="1" rowspan="1" valign="top">12.753 &#x00b1; 2.328 (4.923 &#x2013; 20.3411)</td>
                            <td align="left" colspan="1" rowspan="1" valign="top">392.6077 &#x00b1; 27.1511 (290.63 &#x2013; 436.824)</td>
                            <td align="left" colspan="1" rowspan="1" valign="top">8801.367 &#x00b1; 1085.165 (4789.044 &#x2013; 10761.93)</td>
                        </tr>
                        <tr>
                            <td align="left" colspan="1" rowspan="1" valign="top">
                                <bold>Hippocampus</bold>
                            </td>
                            <td align="left" colspan="1" rowspan="1" valign="top">12.458 &#x00b1; 2.3733 9 (7.467 &#x2013; 19.458)</td>
                            <td align="left" colspan="1" rowspan="1" valign="top">482.367 &#x00b1; 67.634 (310.857 &#x2013; 846.03)</td>
                            <td align="left" colspan="1" rowspan="1" valign="top">13533.601 &#x00b1; 4334.800 (5442.09 &#x2013; 42442.66)</td>
                        </tr>
                        <tr>
                            <td align="left" colspan="1" rowspan="2" valign="top">
                                <bold>Eye (right and left)</bold>
                            </td>
                            <td align="left" colspan="1" rowspan="1" valign="top">4.743 &#x00b1; 2.0700 (0.4666 &#x2013; 13.7333)</td>
                            <td align="left" colspan="1" rowspan="1" valign="top">617.894 &#x00b1; 188.25 (261.39 &#x2013; 1292.476)</td>
                            <td align="left" colspan="1" rowspan="1" valign="top">24411 &#x00b1; 16620 (3849.9 &#x2013; 95551)</td>
                        </tr>
                        <tr>
                            <td align="left" colspan="1" rowspan="1" valign="top">1.875 &#x00b1; 0.6838 (0.842 &#x2013; 4.004)</td>
                            <td align="left" colspan="1" rowspan="1" valign="top">618.840 &#x00b1; 188.2516 (260.266 &#x2013; 1292.476)</td>
                            <td align="left" colspan="1" rowspan="1" valign="top">24499 &#x00b1; 16729 9 3816 &#x2013; 9555)</td>
                        </tr>
                        <tr>
                            <td align="left" colspan="1" rowspan="1" valign="top">
                                <bold>White matter</bold>
                            </td>
                            <td align="left" colspan="1" rowspan="1" valign="top">13.6 &#x00b1; 2.9 (7.8 &#x2013; 22)</td>
                            <td align="left" colspan="1" rowspan="1" valign="top">505.3 &#x00b1; 56 (320.9 &#x2013; 620.4)</td>
                            <td align="left" colspan="1" rowspan="1" valign="top">14734.5 &#x00b1; 3153 (5800.9 + 25444.2)</td>
                        </tr>
                    </tbody>
                </table>
                <table-wrap-foot>
                    <p>CNR - Contrast to Noise ratio, SNR - Signal to Noise ratio, FOM - Figure of Merit.</p>
                </table-wrap-foot>
            </table-wrap>
            <table-wrap id="T5" orientation="portrait" position="float">
                <label>Table 5. </label>
                <caption>
                    <title>Quantitative image parameter &#x2013; Pelvic scans.</title>
                </caption>
                <table content-type="article-table" frame="hsides">
                    <thead>
                        <tr>
                            <th align="left" colspan="1" rowspan="1" valign="top">Region</th>
                            <th align="left" colspan="1" rowspan="1" valign="top">SNR Mean &#x00b1; SD (Range)</th>
                            <th align="left" colspan="1" rowspan="1" valign="top">CNR Mean &#x00b1; SD (Range)</th>
                            <th align="left" colspan="1" rowspan="1" valign="top">FOM Mean &#x00b1; SD (Range)</th>
                        </tr>
                    </thead>
                    <tbody>
                        <tr>
                            <td align="left" colspan="1" rowspan="2" valign="top">
                                <bold>Kidney (right and left)</bold>
                            </td>
                            <td align="left" colspan="1" rowspan="1" valign="top">11.732 &#x00b1; 5.912 (3.55 &#x2013; 27.80)</td>
                            <td align="left" colspan="1" rowspan="1" valign="top">490.19 &#x00b1; 71.83 (328.54 &#x2013; 576.57)</td>
                            <td align="left" colspan="1" rowspan="1" valign="top">13817 &#x00b1; 3632 (6078 &#x2013; 18718.6)</td>
                        </tr>
                        <tr>
                            <td align="left" colspan="1" rowspan="1" valign="top">12.054 &#x00b1; 7.65 (1.88 &#x2013; 38.25)</td>
                            <td align="left" colspan="1" rowspan="1" valign="top">490.19 &#x00b1; 71.83 (330.03 &#x2013; 576.26)</td>
                            <td align="left" colspan="1" rowspan="1" valign="top">13802 &#x00b1; 3601 (6132 &#x2013; 18698)</td>
                        </tr>
                        <tr>
                            <td align="left" colspan="1" rowspan="1" valign="top">
                                <bold>Femur bone</bold>
                            </td>
                            <td align="left" colspan="1" rowspan="1" valign="top">16.21 &#x00b1; 5.46 (4.14 &#x2013; 30.19)</td>
                            <td align="left" colspan="1" rowspan="1" valign="top">533.289 &#x00b1; 26.88 (447.67 &#x2013; 594.15)</td>
                            <td align="left" colspan="1" rowspan="1" valign="top">16113.4 &#x00b1; 1581 (11311 &#x2013; 19889.4)</td>
                        </tr>
                        <tr>
                            <td align="left" colspan="1" rowspan="1" valign="top">
                                <bold>Bladder</bold>
                            </td>
                            <td align="left" colspan="1" rowspan="1" valign="top">2.12 &#x00b1; 1.56 (0.163 &#x2013; 6.683)</td>
                            <td align="left" colspan="1" rowspan="1" valign="top">517.9 &#x00b1; 31.85 (465.3 &#x2013; 577.4)</td>
                            <td align="left" colspan="1" rowspan="1" valign="top">141006 &#x00b1; 26375 (100938 &#x2013; 192500)</td>
                        </tr>
                        <tr>
                            <td align="left" colspan="1" rowspan="1" valign="top">
                                <bold>Fat</bold>
                            </td>
                            <td align="left" colspan="1" rowspan="1" valign="top">-31.48 &#x00b1; 6.82 (-55.43 &#x2013; -13.92)</td>
                            <td align="left" colspan="1" rowspan="1" valign="top">497.97 &#x00b1; 37.32 (421.6 &#x2013; 565.17)</td>
                            <td align="left" colspan="1" rowspan="1" valign="top">14109 &#x00b1; 2087 (10010.3 &#x2013; 18059.1)</td>
                        </tr>
                        <tr>
                            <td align="left" colspan="1" rowspan="1" valign="top">
                                <bold>Muscle</bold>
                            </td>
                            <td align="left" colspan="1" rowspan="1" valign="top">9.22 &#x00b1; 3.4 (3.55 &#x2013; 24.20)</td>
                            <td align="left" colspan="1" rowspan="1" valign="top">531.15 &#x00b1; 35.22 (420.80 &#x2013; 593.55)</td>
                            <td align="left" colspan="1" rowspan="1" valign="top">16024 &#x00b1; 2051 (10034 &#x2013; 19846)</td>
                        </tr>
                    </tbody>
                </table>
                <table-wrap-foot>
                    <p>CNR - Contrast to Noise ratio, SNR - Signal to Noise ratio, FOM - Figure of Merit.</p>
                </table-wrap-foot>
            </table-wrap>
            <p>
                <bold>Determination of optimal protocol on phantom data</bold>
            </p>
            <p>The summary of the technical parameters selected from the experimental study to be used in the optimization process is presented in 
                <xref ref-type="table" rid="T6">Table 6</xref>.</p>
            <table-wrap id="T6" orientation="portrait" position="float">
                <label>Table 6. </label>
                <caption>
                    <title>Quantitative image parameter &#x2013; Phantom study.</title>
                </caption>
                <table content-type="article-table" frame="hsides">
                    <thead>
                        <tr>
                            <th align="left" colspan="5" rowspan="1" valign="top">Phantom data</th>
                            <th align="left" colspan="2" rowspan="1" valign="top">Low-density bone equivalent material</th>
                            <th align="left" colspan="2" rowspan="1" valign="top">High-density bone equivalent material</th>
                        </tr>
                        <tr>
                            <th align="left" colspan="1" rowspan="1" valign="top">kVp</th>
                            <th align="left" colspan="1" rowspan="1" valign="top">mAs</th>
                            <th align="left" colspan="1" rowspan="1" valign="top">CTDI
                                <sub>vol</sub> (mGy)</th>
                            <th align="left" colspan="1" rowspan="1" valign="top">% Reduction</th>
                            <th align="left" colspan="1" rowspan="1" valign="top">DLP</th>
                            <th align="left" colspan="1" rowspan="1" valign="top">SNR</th>
                            <th align="left" colspan="1" rowspan="1" valign="top">CNR</th>
                            <th align="left" colspan="1" rowspan="1" valign="top">SNR</th>
                            <th align="left" colspan="1" rowspan="1" valign="top">CNR</th>
                        </tr>
                    </thead>
                    <tbody>
                        <tr>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">90</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">250</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">6.52</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">63.29</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">205.5</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">43.73</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">266.18</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">152.7</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">371.1</td>
                        </tr>
                        <tr>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">90</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">300</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">7.83</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">55.91</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">222.6</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">43.72</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">249.21</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">172.69</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">346.84</td>
                        </tr>
                        <tr>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">90</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">350</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">9.13</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">48.59</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">259.7</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">45.94</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">258.76</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">190.22</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">376.55</td>
                        </tr>
                        <tr>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">120</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">150</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">8.88</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">50.00</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">252.5</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">46.07</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">259.26</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">170.62</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">356.29</td>
                        </tr>
                        <tr>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">120</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">200</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">11.84</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">33.33</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">336.7</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">44.71</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">244.12</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">168.44</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">379.39</td>
                        </tr>
                        <tr>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">120</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">250</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">14.8</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">16.67</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">420.9</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">47.19</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">266.84</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">160.96</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">378.08</td>
                        </tr>
                        <tr>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">120</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">300</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">17.76</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">0.00</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">505.1</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">51.01</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">280.6</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">200.05</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">399.58</td>
                        </tr>
                        <tr>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">140</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">100</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">8.72</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">50.90</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">248</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">45.06</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">266.29</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">185.88</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">374.33</td>
                        </tr>
                        <tr>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">140</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">150</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">13.08</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">26.35</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">372</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">48.55</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">270.04</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">198.2</td>
                            <td align="left" colspan="1" rowspan="1" valign="bottom">384.93</td>
                        </tr>
                    </tbody>
                </table>
                <table-wrap-foot>
                    <p>kVp - kilovoltage peak, mAs - milli-ampere-second, DLP - Dose Length Product, DRL - Diagnostic reference level, CTDI
                        <sub>vol</sub> - Volumetric computed tomography dose index, CNR - Contrast to Noise ratio, SNR - Signal to Noise ratio.</p>
                </table-wrap-foot>
            </table-wrap>
            <p>The CNR, SNR, and FOM are calculated for ROI created in the phantom scan. The CNR and SNR are compared with the CNR, and SNR is calculated in the patient scans taken in 120 kVp and 300 mAs, selected some combinations of kVp and mAs which was nearer to the patient scan.</p>
        </sec>
        <sec id="sec9" sec-type="discussion">
            <title>4. Discussion</title>
            <p>In the imaging techniques that use ionizing radiation, involving radiotherapy planning computed tomography scans, to minimize patients&#x2019; risk corresponding to the linear no-threshold model, radiation dose needs to be optimized. Diagnostic reference levels are indicators of the typical practice in a country or a region. Because equipment and procedure protocols can vary between different facilities in countries or areas, it is good to establish national or regional diagnostic reference levels.
                <sup>
                    <xref ref-type="bibr" rid="ref14">14</xref>
                </sup> The use of DRL is recognized as a dose optimization tool by many professional and regulatory organizations, including ICRP, ACR, IAEA, and AAPM.</p>
            <p>The focus of the current research was to establish a diagnostic reference level for present RTCT. This study is the first of its kind RTCT simulation DRLs in India. The Radiotherapy CT scanner and the CT scanner used in radiology have some differences. The CT scanner used for the Radiotherapy scan preferably needs to have a big bore size compared to the radiology CT bore. The reason is to allow some specific fixtures used in radiotherapy CT simulation, which is required to maintain the same position of the patient throughout the radiotherapy treatment and avoid some normal structures from the path of radiation. The literature reports a 10 &#x2013; 20 mGy more dose delivered with big bore CT than with small-bore size CT scanner because of the variations in mAs settings.
                <sup>
                    <xref ref-type="bibr" rid="ref15">15</xref>
                </sup> Comparing Radiotherapy CT DRL with Radiology CT scan procedure DRL is not appropriate because of several differences between the two, including different scanning lengths, protocols, and the requirement of different quality of images.
                <sup>
                    <xref ref-type="bibr" rid="ref4">4</xref>
                </sup>
                <sup>,</sup>
                <sup>
                    <xref ref-type="bibr" rid="ref16">16</xref>
                </sup>
            </p>
            <p>Clerkin 
                <italic toggle="yes">et al.</italic>
                <sup>
                    <xref ref-type="bibr" rid="ref17">17</xref>
                </sup> projected a &#x201c;National Irish DRL of 882 mGy cm and CTDI
                <sub>vol</sub> of 21 mGy for the Head and neck RT localization CT. Our study&#x2019;s results comply with this study&#x2019;s recommendations, with the DLP of 790.65 mGy cm and CTDI
                <sub>vol</sub> of 17.76 mGy&#x201d;.</p>
            <p>Nika Zalokar 
                <italic toggle="yes">et al.</italic>
                <sup>
                    <xref ref-type="bibr" rid="ref1">1</xref>
                </sup> projected DRL of 708.2 mGy cm, 663.1 mGy cm, and CTDI
                <sub>vol</sub> of 17.7 mGy, 18.3 mGy for pelvic RT localization CT with Philips and Siemens CT, respectively. The output measurement for the pelvis scan was from the &#x201c;L3-L4&#x201d; portion to 2 cm under the ischiatic bone. Study results with the Dose length product of 999.7 mGy cm and CTDI
                <sub>vol</sub> of 17.76 mGy, and the scan length is D10-D12 region to the mid femur.</p>
            <p>Toro 
                <italic toggle="yes">et al.</italic>
                <sup>
                    <xref ref-type="bibr" rid="ref18">18</xref>
                </sup> studied patient exposure concentrations in Computed tomography models by comparing only CTDI
                <sub>vol</sub> standards. All CTDI
                <sub>vol</sub> values in our study were lower compared to that study. However, wide variations in CTDI
                <sub>vol</sub> have been found among various CT units.</p>
            <p>The current study has a few limitations. This study calculates SNR, CNR, and FOM. However, the correlation between the dose and quality of the image was not addressed based on BMI and chest circumference. The radiation doses that cancer patients obtain throughout imaging have not been significant because of the treatment doses. The RT CT dose optimization is required because cancer patients might require undergoing many CT scans. In the current study, the variation in CTDI
                <sub>vol</sub>, DLP, and scanning protocol in Head and Neck and Pelvis cancer Radiation therapy localization CT imaging in the departments authorize the establishment of DRLs. It may be necessary to consider BMI-specific modified RT CT protocols, and this area warrants further investigation.</p>
        </sec>
        <sec id="sec10" sec-type="conclusion">
            <title>5. Conclusion</title>
            <p>The first regional radiation therapy Computed tomography simulation DRLs have been projected and deliver a platform for dose assessment and optimization. Due to the limited number of literature papers on radiation therapy, Computed tomography DRLs. These provide a basis for dose optimization among RT centres and may facilitate the reduction in cumulative radiation exposure. Comparison with previously established RT CT DRLs showed some radiation dose variation, so exposure parameters should be reviewed and optimized.</p>
        </sec>
        <sec id="sec11">
            <title>Data availability</title>
            <sec id="sec12">
                <title>Underlying data</title>
                <p>Harvard Dataverse, V1. Establishment of Diagnostic reference level and Radiation dose variation in Head &amp; Neck and Pelvis treatment planning in RT CT, 
                    <ext-link ext-link-type="uri" xlink:href="https://doi.org/10.7910/DVN/GIGDBG">https://doi.org/10.7910/DVN/GIGDBG</ext-link>.
                    <sup>
                        <xref ref-type="bibr" rid="ref19">19</xref>
                    </sup>
                </p>
                <p>This project contains the following underlying data: Slice number of the exact location where we had to draw a Region of Interest (ROI) for obtaining the Hounsfield unit and Standard deviation of the area where the ROI is plotted. The collected values from the region of interest will give the Contrast to Noise Ratio (CNR), Signal to Noise Ratio (SNR), and Figure of Merit (FOM). While calculating the radiation dose levels for an individual part, critical parameters like Dose Length Product (DLP), CTDI
                    <sub>vol</sub>, and scan length are considered. All the above data has been liked with the link below.</p>
                <p>Data are available under the terms of the 
                    <ext-link ext-link-type="uri" xlink:href="https://creativecommons.org/publicdomain/zero/1.0/">Creative Commons Zero &#x201c;No rights reserved&#x201d; data waiver</ext-link> (CC0 1.0 Public domain dedication).</p>
            </sec>
        </sec>
    </body>
    <back>
        <ref-list>
            <title>References</title>
            <ref id="ref1">
                <label>1</label>
                <mixed-citation publication-type="journal">
                    <person-group person-group-type="author">

                        <name name-style="western">
                            <surname>Zalokar</surname>
                            <given-names>N</given-names>
                        </name>

                        <name name-style="western">
                            <surname>&#x017d;ager Marciu&#x0161;</surname>
                            <given-names>V</given-names>
                        </name>

                        <name name-style="western">
                            <surname>Meki&#x0161;</surname>
                            <given-names>N</given-names>
                        </name>
</person-group>:
                    <article-title>Establishment of national diagnostic reference levels for radiotherapy computed tomography simulation procedures in Slovenia.</article-title>
                    <source>

                        <italic toggle="yes">Eur. J. Radiol.</italic>
</source>
                    <year>2020 Jun 1</year>;<fpage>127</fpage>.</mixed-citation>
            </ref>
            <ref id="ref2">
                <label>2</label>
                <mixed-citation publication-type="other">
                    <person-group person-group-type="author">

                        <name name-style="western">
                            <surname>McCollough</surname>
                            <given-names>CH</given-names>
                        </name>

                        <name name-style="western">
                            <surname>Clinic</surname>
                            <given-names>M</given-names>
                        </name>
</person-group>:
                    <article-title>Diagnostic Reference Levels.</article-title>
                    <year>2010</year>.</mixed-citation>
            </ref>
            <ref id="ref3">
                <label>3</label>
                <mixed-citation publication-type="journal">
                    <person-group person-group-type="author">

                        <name name-style="western">
                            <surname>Bhosale</surname>
                            <given-names>P</given-names>
                        </name>

                        <name name-style="western">
                            <surname>Wagner-Bartak</surname>
                            <given-names>N</given-names>
                        </name>

                        <name name-style="western">
                            <surname>Wei</surname>
                            <given-names>W</given-names>
                        </name>

                        <etal/>
</person-group>:
                    <article-title>Comparing CNR, SNR, and Image Quality of CT Images Reconstructed with Soft Kernel, Standard Kernel, and Standard Kernel plus ASIR 30% Techniques.</article-title>
                    <source>

                        <italic toggle="yes">Int. J. Radiol.</italic>
</source>
                    <year>2015</year>;<volume>2</volume>(<issue>2</issue>):<fpage>60</fpage>&#x2013;<lpage>65</lpage>.
                    <pub-id pub-id-type="doi">10.17554/j.issn.2313-3406.2015.02.11</pub-id>
                    <ext-link ext-link-type="uri" xlink:href="http://www.ghrnet.org/index.php/ijr/article/view/1101">Reference Source</ext-link>
                </mixed-citation>
            </ref>
            <ref id="ref4">
                <label>4</label>
                <mixed-citation publication-type="book">
                    <person-group person-group-type="author">

                        <name name-style="western">
                            <surname>Mutic</surname>
                            <given-names>S</given-names>
                        </name>

                        <name name-style="western">
                            <surname>Palta</surname>
                            <given-names>JR</given-names>
                        </name>

                        <name name-style="western">
                            <surname>Butker</surname>
                            <given-names>EK</given-names>
                        </name>

                        <etal/>
</person-group>:
                    <source>

                        <italic toggle="yes">Quality assurance for computed-tomography simulators and the computed-tomography-simulation process: Report of the AAPM Radiation Therapy Committee Task Group No. 66. Vol. 30, Medical Physics.</italic>
</source>
                    <publisher-name>John Wiley and Sons Ltd</publisher-name>;<year>2003</year>;<fpage>2762</fpage>&#x2013;<lpage>2792</lpage>.</mixed-citation>
            </ref>
            <ref id="ref5">
                <label>5</label>
                <mixed-citation publication-type="journal">
                    <person-group person-group-type="author">

                        <name name-style="western">
                            <surname>Chang</surname>
                            <given-names>KP</given-names>
                        </name>

                        <name name-style="western">
                            <surname>Hsu</surname>
                            <given-names>TK</given-names>
                        </name>

                        <name name-style="western">
                            <surname>Lin</surname>
                            <given-names>WT</given-names>
                        </name>

                        <etal/>
</person-group>:
                    <article-title>Optimization of dose and image quality in adult and pediatric computed tomography scans.</article-title>
                    <source>

                        <italic toggle="yes">Radiat. Phys. Chem.</italic>
</source>
                    <year>2017 Nov 1</year>;<volume>140</volume>:<fpage>260</fpage>&#x2013;<lpage>265</lpage>.
                    <pub-id pub-id-type="doi">10.1016/j.radphyschem.2017.02.006</pub-id>
                </mixed-citation>
            </ref>
            <ref id="ref6">
                <label>6</label>
                <mixed-citation publication-type="other">
                    <article-title>Pulmonary Nodules Experimental and Clinical Studies at LowDose CT.</article-title>
                </mixed-citation>
            </ref>
            <ref id="ref7">
                <label>7</label>
                <mixed-citation publication-type="journal">
                    <person-group person-group-type="author">

                        <name name-style="western">
                            <surname>Itoh</surname>
                            <given-names>S</given-names>
                        </name>

                        <name name-style="western">
                            <surname>Ikeda</surname>
                            <given-names>M</given-names>
                        </name>

                        <name name-style="western">
                            <surname>Isomura</surname>
                            <given-names>T</given-names>
                        </name>

                        <etal/>
</person-group>:
                    <article-title>Screening helical CT for mass screening of lung cancer: application of low-dose and single-breath-hold scanning.</article-title>
                    <source>

                        <italic toggle="yes">Radiat. Med.</italic>
</source>
                    <year>1998 Mar-Apr</year>;<volume>16</volume>(<issue>2</issue>):<fpage>75</fpage>&#x2013;<lpage>83</lpage>.
                    <pub-id pub-id-type="pmid">9650893</pub-id>
                </mixed-citation>
            </ref>
            <ref id="ref8">
                <label>8</label>
                <mixed-citation publication-type="book">
                    <person-group person-group-type="author">

                        <name name-style="western">
                            <surname>Radiology</surname>
                            <given-names>E</given-names>
                        </name>

                        <name name-style="western">
                            <surname>Gartenschl&#x00e4;ger</surname>
                            <given-names>M</given-names>
                        </name>

                        <name name-style="western">
                            <surname>Schweden</surname>
                            <given-names>F</given-names>
                        </name>

                        <etal/>
</person-group>:
                    <source>

                        <italic toggle="yes">Chest radiology Pulmonary nodules: detection with low-dose vs conventional-dose spiral CT.</italic>
</source>vol.<volume>8</volume>
                    <publisher-name>Springer-Verlag</publisher-name>;<year>1998</year>.</mixed-citation>
            </ref>
            <ref id="ref9">
                <label>9</label>
                <mixed-citation publication-type="journal">
                    <person-group person-group-type="author">

                        <name name-style="western">
                            <surname>Bj&#x00f6;rkdahl</surname>
                            <given-names>P</given-names>
                        </name>

                        <name name-style="western">
                            <surname>Nyman</surname>
                            <given-names>U</given-names>
                        </name>
</person-group>:
                    <article-title>Using 100- instead of 120-kVp computed tomography to diagnose pulmonary embolism almost halves the radiation dose with preserved diagnostic quality.</article-title>
                    <source>

                        <italic toggle="yes">Acta Radiol.</italic>
</source>
                    <year>2010</year>;<volume>51</volume>(<issue>3</issue>):<fpage>260</fpage>&#x2013;<lpage>270</lpage>.
                    <pub-id pub-id-type="pmid">20121673</pub-id>
                    <pub-id pub-id-type="doi">10.3109/02841850903505222</pub-id>
                </mixed-citation>
            </ref>
            <ref id="ref10">
                <label>10</label>
                <mixed-citation publication-type="journal">
                    <person-group person-group-type="author">

                        <name name-style="western">
                            <surname>Heyer</surname>
                            <given-names>CM</given-names>
                        </name>

                        <name name-style="western">
                            <surname>Mohr</surname>
                            <given-names>PS</given-names>
                        </name>

                        <name name-style="western">
                            <surname>Lemburg</surname>
                            <given-names>SP</given-names>
                        </name>

                        <etal/>
</person-group>:
                    <article-title>Image quality and radiation exposure at pulmonary CT angiography with 100- or 120-kVp protocol: Prospective randomized study.</article-title>
                    <source>

                        <italic toggle="yes">Radiology.</italic>
</source>
                    <year>2007 Nov</year>;<volume>245</volume>(<issue>2</issue>):<fpage>577</fpage>&#x2013;<lpage>583</lpage>.
                    <pub-id pub-id-type="doi">10.1148/radiol.2452061919</pub-id>
                </mixed-citation>
            </ref>
            <ref id="ref11">
                <label>11</label>
                <mixed-citation publication-type="journal">
                    <person-group person-group-type="author">

                        <name name-style="western">
                            <surname>Chen</surname>
                            <given-names>GP</given-names>
                        </name>

                        <name name-style="western">
                            <surname>Noid</surname>
                            <given-names>G</given-names>
                        </name>

                        <name name-style="western">
                            <surname>Tai</surname>
                            <given-names>A</given-names>
                        </name>

                        <etal/>
</person-group>:
                    <article-title>Improving CT quality with optimized image parameters for radiation treatment planning and delivery guidance.</article-title>
                    <source>

                        <italic toggle="yes">Phys. Imaging Radiat. Oncol.</italic>
</source>
                    <year>2017 Oct 1</year>;<volume>4</volume>:<fpage>6</fpage>&#x2013;<lpage>11</lpage>.
                    <pub-id pub-id-type="doi">10.1016/j.phro.2017.10.003</pub-id>
                </mixed-citation>
            </ref>
            <ref id="ref12">
                <label>12</label>
                <mixed-citation publication-type="journal">
                    <person-group person-group-type="author">

                        <name name-style="western">
                            <surname>Singh</surname>
                            <given-names>S</given-names>
                        </name>

                        <name name-style="western">
                            <surname>Kalra</surname>
                            <given-names>MK</given-names>
                        </name>

                        <name name-style="western">
                            <surname>Moore</surname>
                            <given-names>MA</given-names>
                        </name>

                        <etal/>
</person-group>:
                    <article-title>Dose reduction and compliance with pediatric CT protocols adapted to patient size, clinical indication, and the number of prior studies.</article-title>
                    <source>

                        <italic toggle="yes">Radiology.</italic>
</source>
                    <year>2009 Jul</year>;<volume>252</volume>(<issue>1</issue>):<fpage>200</fpage>&#x2013;<lpage>208</lpage>.
                    <pub-id pub-id-type="pmid">19435938</pub-id>
                    <pub-id pub-id-type="doi">10.1148/radiol.2521081554</pub-id>
                </mixed-citation>
            </ref>
            <ref id="ref13">
                <label>13</label>
                <mixed-citation publication-type="journal">
                    <person-group person-group-type="author">

                        <name name-style="western">
                            <surname>Singh</surname>
                            <given-names>S</given-names>
                        </name>

                        <name name-style="western">
                            <surname>Kalra</surname>
                            <given-names>MK</given-names>
                        </name>

                        <name name-style="western">
                            <surname>Ali Khawaja</surname>
                            <given-names>RD</given-names>
                        </name>

                        <etal/>
</person-group>:
                    <article-title>Radiation Dose Optimization and Thoracic Computed Tomography.</article-title>
                    <source>

                        <italic toggle="yes">Radiol. Clin. N. Am.</italic>
</source>
                    <year>2014</year>;<volume>52</volume>:<fpage>1</fpage>&#x2013;<lpage>15</lpage>.
                    <pub-id pub-id-type="doi">10.1016/j.rcl.2013.08.004</pub-id>
                </mixed-citation>
            </ref>
            <ref id="ref14">
                <label>14</label>
                <mixed-citation publication-type="journal">
                    <person-group person-group-type="author">

                        <name name-style="western">
                            <surname>Vassileva</surname>
                            <given-names>J</given-names>
                        </name>

                        <name name-style="western">
                            <surname>Rehani</surname>
                            <given-names>M</given-names>
                        </name>
</person-group>:
                    <article-title>Diagnostic reference levels.</article-title>
                    <source>

                        <italic toggle="yes">AJR Am. J. Roentgenol.</italic>
</source>
                    <year>2015 Jan 1</year>;<volume>204</volume>(<issue>1</issue>):<fpage>W1</fpage>&#x2013;<lpage>W3</lpage>.
                    <pub-id pub-id-type="doi">10.2214/AJR.14.12794</pub-id>
                </mixed-citation>
            </ref>
            <ref id="ref15">
                <label>15</label>
                <mixed-citation publication-type="journal">
                    <person-group person-group-type="author">

                        <name name-style="western">
                            <surname>Garcia-Ramirez</surname>
                            <given-names>JL</given-names>
                        </name>

                        <name name-style="western">
                            <surname>Dempsey</surname>
                            <given-names>JF</given-names>
                        </name>

                        <name name-style="western">
                            <surname>Low</surname>
                            <given-names>DA</given-names>
                        </name>

                        <etal/>
</person-group>:
                    <article-title>Performance evaluation of an 85-cm-bore x-ray computed tomography scanner designed for radiation oncology and comparison with current diagnostic CT scanners.</article-title>
                    <source>

                        <italic toggle="yes">Int. J. Radiat. Oncol. Biol. Phys.</italic>
</source>
                    <year>2002</year>;<volume>52</volume>:<fpage>1123</fpage>&#x2013;<lpage>1131</lpage>.
                    <pub-id pub-id-type="pmid">11958910</pub-id>
                    <pub-id pub-id-type="doi">10.1016/S0360-3016(01)02779-1</pub-id>
                </mixed-citation>
            </ref>
            <ref id="ref16">
                <label>16</label>
                <mixed-citation publication-type="other">
                    <person-group person-group-type="author">

                        <name name-style="western">
                            <surname>Mc Ardle</surname>
                            <given-names>O</given-names>
                        </name>

                        <name name-style="western">
                            <surname>Mullaney</surname>
                            <given-names>L</given-names>
                        </name>
</person-group>:
                    <article-title>Title: Establishment of national diagnostic reference levels for breast cancer CT protocols in radiation therapy. Running Title: Establishment of DRLs for breast cancer CT protocols in RT.</article-title>
                    <year>2016</year>.</mixed-citation>
            </ref>
            <ref id="ref17">
                <label>17</label>
                <mixed-citation publication-type="journal">
                    <person-group person-group-type="author">

                        <name name-style="western">
                            <surname>Clerkin</surname>
                            <given-names>C</given-names>
                        </name>

                        <name name-style="western">
                            <surname>Brennan</surname>
                            <given-names>S</given-names>
                        </name>

                        <name name-style="western">
                            <surname>Mullaney</surname>
                            <given-names>LM</given-names>
                        </name>
</person-group>:
                    <article-title>Establishment of national diagnostic reference levels (DRLs) for radiotherapy localization computer tomography of the Head and neck.</article-title>
                    <source>

                        <italic toggle="yes">Rep. Pract. Oncol. Radiother.</italic>
</source>
                    <year>2018 Sep 1</year>;<volume>23</volume>(<issue>5</issue>):<fpage>407</fpage>&#x2013;<lpage>412</lpage>.
                    <pub-id pub-id-type="pmid">30190653</pub-id>
                    <pub-id pub-id-type="doi">10.1016/j.rpor.2018.07.012</pub-id>
                </mixed-citation>
            </ref>
            <ref id="ref18">
                <label>18</label>
                <mixed-citation publication-type="journal">
                    <person-group person-group-type="author">

                        <name name-style="western">
                            <surname>Toroi</surname>
                            <given-names>P</given-names>
                        </name>

                        <name name-style="western">
                            <surname>Kaijaluoto</surname>
                            <given-names>S</given-names>
                        </name>

                        <name name-style="western">
                            <surname>Bly</surname>
                            <given-names>R</given-names>
                        </name>
</person-group>:
                    <article-title>Patient exposure levels in radiotherapy CT simulations in Finland.</article-title>
                    <source>

                        <italic toggle="yes">Radiat. Prot. Dosim.</italic>
</source>
                    <year>2015 Dec 1</year>;<volume>167</volume>(<issue>4</issue>):<fpage>602</fpage>&#x2013;<lpage>607</lpage>.
                    <pub-id pub-id-type="pmid">25543133</pub-id>
                    <pub-id pub-id-type="doi">10.1093/rpd/ncu363</pub-id>
                </mixed-citation>
            </ref>
            <ref id="ref19">
                <label>19</label>
                <mixed-citation publication-type="other">
                    <person-group person-group-type="author">

                        <name name-style="western">
                            <surname>Suresh</surname>
                            <given-names>S</given-names>
                        </name>
</person-group>:
                    <article-title>Establishment of Diagnostic reference level and Radiation dose variation in Head &amp; Neck and Pelvis treatment planning in RTCT.</article-title>
                    <source>

                        <italic toggle="yes">Harvard Dataverse, V1.</italic>
</source>
                    <year>2022</year>.
                    <pub-id pub-id-type="doi">10.7910/DVN/GIGDBG</pub-id>
                </mixed-citation>
            </ref>
        </ref-list>
    </back>
</article>