Brief Report
Revised

Coiling of echogenic perineural catheters with integral stylet: A proof-of-concept randomized control trial in a sciatic nerve block simulator and a pilot study in orthopaedic-trauma patients.

[version 2; peer review: 2 approved]
PUBLISHED 02 Dec 2024
Author details Author details
OPEN PEER REVIEW
REVIEWER STATUS

Abstract

Backround/Objectives

We investigated a technique that facilitates the coiling of a regular straight catheter (with integral stylet) behind the sciatic nerve in an ultrasound (US) regional anaesthesia simulator, and then applied our findings to a series of orthopedic-trauma patients.

Methods

We conducted a randomized study of two methods of perineural catheter advancement in a sciatic nerve block Blue Phantom simulator. Two groups of twenty catheters each (method A and method B) were evaluated under real-time ultrasound imaging. The needle in-plane/nerve in-short-axis technique was applied. In method A the catheter was advanced beyond the needle tip with the integral stylet extending along its entire length; in method B the catheter was advanced after its integral stylet was retracted by 6 cm, thus providing flexibility to the catheter’s distal end. Additionally, to assess the procedural effectiveness of method B coiling technique, a pilot study was conducted examining 25 perineural catheters coiled underneath the sciatic nerve in trauma-orthopaedic patients to document any catheter tip displacement from their initial position (for 36 hours postoperatively).

Results

In the simulation study, method B led to a significantly higher percentage (18/20:90%) of coiled catheters than method A (3/20:15%). Two coiled catheters of method B were found kinked/obstructed. In our patients, after catheter insertion, the distal end of 2/25 (8%) coiled catheters was obstructed. One perineural catheter was dislodged. For the remaining 22 (88%) catheters, ultrasound imaging demonstrated that local anaesthetic infusion made contact with the sciatic nerve, indicating no displacement of the catheter’s distal end postoperatively.

Conclusion

Regular straight perineural catheters can be coiled if their integral stylet is partially retracted. This coiling method offers extra catheter length adjacent to the nerve structure which potentially mitigates catheter tip displacement.

Trial registration

clinicaltrials.gov, registration No: NCT06568510, 23/08/2024, registration URL: https://clinicaltrials.gov/study/NCT06568510?intr=coiling%20of%20echogenic%20sciatic%20nerve&rank=1#study-overview

Keywords

perineural catheter coiling; sciatic nerve block; ultrasound; catheter tip dislocation; simulation.

Revised Amendments from Version 1

Differences between this version of the article and the previously published version

1. Purpose of the study: the second paragraph of the introduction section was redrafted, further and better elaborating on the purpose of the study. The reference of Nickl R et all is now cited in the reference list for the same reason.

2. Coiling manoeuvre video: the coiling manoeuver of the catheter was recognizable due to its dynamic coiling motion during advancement (frame by frame analysis of the catheter in the stored video loops). A video clearly presenting the motion of the catheter and its coiling manoeuver during advancement was added.

3. Coiling technique: The technique is more clearly described by incorporating in the text more details regarding catheter’s placement in the paraneurium

4. Pilot study: our clinical study was not powered, nor did it include a control group. This is an obvious methodology weakness, even though the intention was to deploy a pilot study supplementary to the simulation study, in addition to providing definitive preliminary data for a further randomized control clinical trial. We concur with the reviewer’s comment and a new paragraph in the limitation section is thus added.

5. Figure 3: Figure 3 is omitted as proposed by reviewer 2, and an explanatory video is included (data availability section).

See the authors' detailed response to the review by Konstantinos Kalimeris
See the authors' detailed response to the review by Agathi Karakosta

Trial registration

clinicaltrials.gov, registration No: NCT06568510, 23/08/2024, registration URL: https://clinicaltrials.gov/study/NCT06568510?intr=coiling%20of%20echogenic%20sciatic%20nerve&rank=1#study-overview

Introduction

Self-coiling catheters and catheters with a flexible, distal end-portion and wire-reinforced body can be looped, providing additional catheter length adjacent to the targeted nerves. Coiling potentially lowers dislocation rates of continuous peripheral nerve block (CPNB) catheters.1,2

Regular straight perineural catheters inserted with the needle in-plane/nerve in-short-axis technique, tend to bypass the nerve due to the perpendicular orientation of the needle/catheter relative to the nerve.3,4 Furthermore, they demonstrate high in situ dislocation rates and subsequently CPNB failure.1 As of this time, there are no investigations examining whether regular straight perineural catheters (inserted with the needle in-plane/nerve in-short-axis technique) can be looped adjacent to nerves, and consequently, mitigate CPNB failure secondary to displacement.

Therefore, we investigated a technique that facilitates the coiling of a regular straight catheter with integral stylet (polyurethane based) behind the sciatic nerve in an ultrasound (US) regional anaesthesia simulator, and then applied our findings to a series of orthopedic-trauma patients.

Methods

Study design and settings

Two independent studies were conducted: 1) A simulation-based protocol in a Blue Phantom US simulator of sciatic nerve block. 2) A pilot study in 25 patients undergoing sciatic CPNB. All procedures were performed under US real-time visualization (5-9 MHz linear transducer; LOGIQ e; GE Healthcare, USA).

Simulation-based CPNBs methods and randomization

Two methods of perineural catheter advancement using the Blue Phantom simulator were evaluated. A 20G single orifice, echogenic straight perineural catheter, polyurethane based (communication with the manufacturer) with a stainless-steel helical coil-reinforced body and a steel integral stylet (SonoLong/Sono, PAJUNK, Germany) was used.

Method A: The catheter was advanced beyond the needle tip (under real-time US imaging) with its integral stylet extending along the entire length of the catheter ( Figure 1A).

6a076808-a7dc-423f-a7c9-e928f9675b29_figure1.gif

Figure 1. A: The perineural catheter removed from catheter’s container, with the integral stylet extending along its entire length (from the beginning to the tip); B: The catheter’s integral stylet is retracted by 6 cm (arrow) so that the catheter’s distal portion features a more flexible end; C: Blue Phantom’s sciatic nerve in long axis view; D: the needle insertion angle (40 degrees) to the Blue Phantom posterior thigh surface as measured by goniometer; E: Ultrasound image of the Tuohy needle tip (arrow) as directed underneath the sciatic nerve of the simulator.

Inset: perineural catheter container. SC.N=Sciatic Nerve.

Method B: The catheter was advanced beyond the needle tip after the catheter’s integral stylet was retracted by 6 cm so that the catheter’s distal portion provided a more flexible end ( Figure 1B).

Randomization was done using a randomization generator (www.randomization.com) by means of sequentially numbered sealed opaque envelopes, including information about the sequence of method allocation.

The simulator was provided with a sciatic nerve insert (upper third of the posterior thigh) containing one nerve structure (CAE training Blue Phantom). The sciatic nerve-simulator surface distance featured different location depths (2-4 cm) ( Figure 1C).

A Tuohy needle (18G × 7.5 mm, PAJUNK, Germany) was inserted at the simulator’s posterior thigh (TS) from a lateral to medial direction (needle in-plane/nerve in-short-axis technique). The needle insertion angles (coronal plane) were always kept at 40 degrees ( Figure 1D). In both methods, the needle was aimed behind the sciatic nerve with the curved end facing the nerve ( Figure 1E). The perineural catheter was threaded through the Tuohy needle until its tip reached the needle orifice. The catheter was slowly advanced exactly 6 cm (under real-time US guidance) beyond the curved tip of the Tuohy needle, until the catheter’s distal end-portion coiled exactly under the sciatic nerve ( Figure 2A). To better visualize a catheter’s trajectory, a slight heel-toe maneuver was implemented to bring the US beam perpendicular to catheter’s distal portion. Procedural time was limited to 15 seconds (total allowed performance time for each coiling maneuver). Each attempt was stored digitally5 and analyzed off-line (frame by frame) by two co-investigators (EP, MR) who were blinded to group allocation. The two investigators decided independently whether the catheter coiled or not. The result was considered positive (coiled catheter) only after unanimous agreement between the two investigators. If the catheter distal end did not coil or if the catheter bypassed the nerve and then coiled, the catheter was withdrawn, the needle tip reoriented (rotation of needle shaft/tip at 90 degrees), and the catheter was readvanced. In total, three attempts were allowed (overall maximum procedural time: 45 seconds). The number of attempts to coil a catheter, as well as the proportion of cases where a catheter was successfully looped on first attempt, were measured. The patency of coiled catheters was tested after normal saline (3 ml) infusion through the catheter. Contact of saline spread with the nerve was also recorded ( Figure 2B). All images were taken under the same imaging settings (gain: 90 db, focus position at the level of the nerve, standard time gain compensation).

6a076808-a7dc-423f-a7c9-e928f9675b29_figure2.gif

Figure 2. A: The coiled perineural catheter underneath the sciatic nerve (arrowheads) of the Blue Phantom simulator; B: Saline spread contacts the sciatic nerve (arrowheads) of the Blue Phantom simulator; C: The injection of LA through the perineural catheter (arrows) contacts the pantient’s sciatic nerve (arrowheads) prior to surgery; and D: 36 hours postoperatively.

C=Perineural Catheter, SC.N=Sciatic Nerve, *=LA spread.

Statistical analysis

Sample size calculation was based on a previous pilot simulation study including 10 CPNBs for each method, using a Blue Phantom nerve block simulator that contained two nerve structures with dimensions of 17cm × 13cm × 6cm (L × H × W), (CAE-Healthcare). Catheters were coiled (after maximum of three attempts) in 3/10 (30%) and in 8/10 (80%) cases, with method A and B respectively. Using an α error of 0.025 and a power of (1− β) at 80% to detect a 50% difference between the two methods, a sample size of 18 CPNBs in each group was required. To allow for increased data variability, 40 CPNBs were performed in a sciatic nerve block simulator. The D’Agostino-Pearson test assessed normality. The results were expressed as percentage (%) or median (highest value-lowest value). Proportions were compared using Fisher’s exact test; p values <0.05 were considered statistically significant. Interobserver agreement was assessed by using the inter-rater agreement statistic Kappa (MedCalc Software, Mariakerke, Belgium).

Pilot study in orthopaedic-trauma patients

To assess procedural effectiveness of the preferred coiling technique (method B), a pilot study was performed. The pilot study was conducted according to the principles expressed in the Declaration of Helsinki and was approved by the “Attikon” Hospital Ethics Committee (Approval No 342, Approval Date 09/05/2024). Informed written consent was always obtained from the patients or their surrogates.

Straight perineural catheters with coiled distal ends behind the sciatic nerve were examined in 25 patients (age: 28-55 years; male/female: 14/11; BMI: 26 (19-31) kg/m2) who had sustained tibia fractures (recruitment period: 10/05/2024-15/07/2024). Exclusion criteria: patients with infection in the subgluteal region, allergies to local anaesthetics, patients’ refusal to consent for the nerve block, neurological disorders, pregnancy.

Patients were placed in the lateral position (with the injured leg uppermost) and a US-guided sciatic nerve block was performed under aseptic conditions (TS, EP). The sciatic nerve was visualized at the upper third of the posterior thigh (short axis view). A Tuohy needle (18 G × 10 mm, SonoLong/NanoLine, PAJUNK, Germany) was inserted in a lateral to medial direction. The needle was gently advanced into the subparaneural compartment (area between the paraneural sheath and epineurium). An initial 3 ml normal saline injection distinctly elevated the various layers of the paraneural tissues and dynamically shifted the sciatic nerve upwards, thus confirming proper needle tip placement in the paraneural compartment.

Ten ml of ropivacaine (0.1%) were then injected and the LA spread diffused in this space between the multiple layers of the paraneural tissues without nerve expansion (thus excluding intraneural injection). A perineural catheter (with retracted integral stylet) (SonoLong / Sono, PAJUNK, Germany) was threaded through the needle tip and coiled behind the sciatic nerve in the paraneural space. Specifically, confirmation of correct catheter tip placement in the peraneural space was defined as ropivacaine injectate (3 ml, 0.1%) spread was distributed around and in contact with the sciatic nerve ( Figure 2C). If the distribution of the LA could not be visualized on the first infusion, 2 additional injections were performed until the distribution was clearly seen. If the LA spread could not be visually confirmed, or if, when confirmed, LA was not distributed around and in contact with the sciatic nerve, the case was excluded.

Subcutaneous tunneling (4 cm long) and placement of a transparent adhesive dressing were used to secure the catheter. All sciatic nerve blocks were followed by an adductor canal block (10 ml of ropivacaine 0.5%) and general anaesthesia. A continuous infusion 8-12 ml of ropivacaine 0,1% (ROPIVACAINE/KABI INJ.SOL 2 mg/ml, 28823.01.08, Fresenius Kabi Hellas A.E.) was commenced and an oral combination of paracetamol 325 mg with oxycodone 5 mg (DEPALGOS F.C. TAB 325+5 mg, 91337.01.01, MOLTENI SPA, Italy) was provided on an as-needed basis after surgery.

All CPNBs were assessed and data were collected in the post-anaesthetic care unit, and every 12 hours thereafter until 36 hours postoperatively from members of the acute pain service (APS) team not participating in the study. Observation of fluid under the dressing indicated catheter’s leakage. Unplanned external displacement and dislodgement of the catheter were defined: less than and equal to 6 cm (displacement) and more than 6 cm (dislodgement) movement of the catheter from the initial recorded depth at the point of insertion or from the distal end of the subcutaneous tunnel. The final catheter tip position was evaluated (under US imaging) 36 hours postoperatively and was identified by injecting 3 ml of ropivacaine (0.1%) through the catheter ( Figure 2D). If LA spread was not confirmed after 3 injections or if, when confirmed, it did not come in contact with and around the sciatic nerve, the catheter’s tip was considered dislocated.

Results

In the simulation study, overall, 21 catheters coiled behind the sciatic nerve in both groups. None of the catheters coiled after bypassing the nerve. Removal of integral stylet (group B) led to a higher percentage (18/20:90%) of coiling catheters than in group A (3/20:15%), p=0.0003. Of group A (n=3) and group B (n=18) coiled catheters, 3/3 (100%) and 15/18 (83%) (p=0.9) coiled on first attempt, respectively. There was significant agreement (kappa=0.9) between the two investigators assessing perineural catheter coiling. In 2 out of 18 coiled catheters (group B), saline injection was not feasible due to catheter’s distal end kinking/obstruction. Saline injections contacted the nerve in 3/3 (100%) (group A) and in 13/16 (81%) (group B) coiled catheters, respectively. After gradual uncoiling of the 2 obstructed catheters by 1 and 2 cm respectively, successful injection of saline became possible and made contact with the nerve.6

Neither catheter knotting nor catheter shear/damage were observed when the catheter was retracted and readvanced to attain proper target placement.

In the pilot study, the distal end of 2/25 (8%) coiled catheters was kinked/obstructed after catheter insertion, but after uncoiling the catheters by 3 cm and 2 cm, ropivacaine administration was successful. One catheter was dislodged postoperatively. Of the 22 indwelling catheters evaluated 36 hours postoperatively, the LA spread was visualized in all US examinations. Although 5/22 (22%) catheters were found externally displaced (displacement length: 2 (2-3) cm) postoperatively, no catheter tips were found dislocated. No leakage around the catheters was noticed. Neither catheter knotting nor nerve trapping/injury were recorded. Oxycodone consumption was 12.5 (0-20) mg and the average numeric rating scale (NRS) score: 2 (0-5).7

Discussion

A novel methodology to sufficiently evaluate needles and perineural catheters on patients (not following manufacturer’s instructions) entails potential risks.812 Therefore, we decided to initially assess this specific perineural catheter method in a “human-tissue mimicking” simulation model.

The Blue Phantom simulator was selected because its physical properties are close to human tissue and has been employed as a testing method in peripheral nerve block needle visibility research.1318 Additionally, the Blue Phantom sciatic nerve block simulator exhibits an anatomic configuration approximating that of human anatomy.

Furthermore, by partially withdrawing the integral stylet, the catheter body remained resistant to bending, while the catheter’s distal end developed a balance of malleability and strength that was required to achieve an ease of coiling, as well as lower probability for inadvertent penetration and nerve injury. We also used a Tuohy needle because its curved tip facilitated a trajectory (directional curvature) of the catheter that allowed proper target placement.

In our simulation study, the needle-insertion angles were maintained constant to attain standard angulations between the catheter’s tip and the surrounding Blue Phantom material once the catheter was threaded out of the Tuohy needle. We did not use shallow 0 to 30 degrees catheter insertion angles to optimize echogenicity because the Blue Phantom’s material is homogenous with low background echogenicity and enhances needle (and consequently perineural catheter) visibility, especially at low needle insertion angles.8 We decided, thus, to implement a steeper insertion angle of 40 degrees to simulate more difficult clinical conditions. Nevertheless, in all cases, the operator could easily advance the catheter and the two independent investigators could visually assess the trajectory of the catheters with precision and significant interobserver agreement.

In our patients, US examination of the perineural catheters (36 hours postoperatively) did not reveal any catheter tip dislocation (internally migrated catheter tip) from their initial position placement (underneath the sciatic nerve); even though 22% (5/22) of them exhibited external displacement at the point of insertion or from the distal end of the subcutaneous tunnel. It has been previously reported that regular straight perineural catheters with integral stylet display a lower tip dislocation rate when placed with the needle out-of-plane/nerve in-short axis approach.4,19 However, we can advocate that this type of perineural catheter could alternatively be inserted with good results, i.e. low dislocation rates of catheter’s tip, if the needle in-plane/nerve in-short axis approach is combined with the coiling CPNB technique.

Leakage of perineural catheters is a shortcoming of the catheter-through-needle method.20 In our case series, no perineural leakage was recorded. It could be considered that a 6 cm coiled catheter path in perineural tissues may be long and winding enough to prevent local anesthetic from flowing back through the needle-puncture track. However, further investigation is imperative to support this hypothesis.

A procedural weakness is the kinking of the coiled catheters’ distal end. Nevertheless, by partially uncoiling the catheter, points of obstruction can be relieved. We based our 6 cm stylet retraction on a previous report which although using a different perineural catheter brand, had a fixed 6 cm flexible/soft distal end as well.2 Although an advantage of this technique is that the stylet can be retracted from the catheter’s tip at different lengths, we did not study various lengths of stylet withdrawal (and consequently the optimal length affecting a catheter’s flexible end) that could facilitate the coiling maneuver without obstruction of the catheter.

A shortcoming of our study was that the coiling ability of this type of perineural catheters was not clinically examined; however, this question had been answered through the simulation study. Additionally, the examination of a catheter’s displacement from its initial position (behind the sciatic nerve) in a Blue Phantom simulator is not feasible. Of note, our clinical study was not powered, nor did it include a control group. In fact, our pilot study was complementary to our simulation study, and its results need to be strengthened by a future randomized control clinical trial.

Moreover, attention should be paid to if extrapolating these observations to other ultrasound approaches, perineural catheter placement methods, as well as to diverse time-intervals of LA infusions.

Conclusion

A regular straight perineural catheter can be coiled if its integral stylet is partially retracted. This coiling method secures a safe and adequate extra length that delivers to the operator a margin of comfort regarding the mitigation of perineural catheter tip dislocation. Further studies are required to evaluate the clinical effectiveness of this perineural catheter coiling method over other catheters’ designs and techniques.

Ethics and consent

The pilot study was conducted according to the principles expressed in the Declaration of Helsinki and was approved by the “Attikon” Hospital Ethics Committee (Approval No 342, Approval Date 09/05/2024). In accordance with the “Attikon” Hospital Ethics Committee requirements, informed written consent was always obtained from the patients or their surrogates.

Author contributions

TS: conception and implementation of the technique, writing, and correcting the manuscript. EP: equal contribution to the first author at all stages of the study. MR and KP: data collection and analysis. AM: recruiment and follow-up of patients. TP: commenting on all stages of the study and drafting the manuscript. All authors read and approved this final version of the manuscript.

Comments on this article Comments (0)

Version 2
VERSION 2 PUBLISHED 27 Sep 2024
Comment
Author details Author details
Competing interests
Grant information
Copyright
Download
 
Export To
metrics
Views Downloads
F1000Research - -
PubMed Central
Data from PMC are received and updated monthly.
- -
Citations
CITE
how to cite this article
Saranteas T, Poulogiannopoulou E, Riga M et al. Coiling of echogenic perineural catheters with integral stylet: A proof-of-concept randomized control trial in a sciatic nerve block simulator and a pilot study in orthopaedic-trauma patients. [version 2; peer review: 2 approved]. F1000Research 2024, 13:1103 (https://doi.org/10.12688/f1000research.155381.2)
NOTE: If applicable, it is important to ensure the information in square brackets after the title is included in all citations of this article.
track
receive updates on this article
Track an article to receive email alerts on any updates to this article.

Open Peer Review

Current Reviewer Status: ?
Key to Reviewer Statuses VIEW
ApprovedThe paper is scientifically sound in its current form and only minor, if any, improvements are suggested
Approved with reservations A number of small changes, sometimes more significant revisions are required to address specific details and improve the papers academic merit.
Not approvedFundamental flaws in the paper seriously undermine the findings and conclusions
Version 2
VERSION 2
PUBLISHED 02 Dec 2024
Revised
Views
2
Cite
Reviewer Report 10 Dec 2024
Agathi Karakosta, University of Ioannina, Ioannina, Greece 
Approved
VIEWS 2
I have read the revised manuscript carefully. ... Continue reading
CITE
CITE
HOW TO CITE THIS REPORT
Karakosta A. Reviewer Report For: Coiling of echogenic perineural catheters with integral stylet: A proof-of-concept randomized control trial in a sciatic nerve block simulator and a pilot study in orthopaedic-trauma patients. [version 2; peer review: 2 approved]. F1000Research 2024, 13:1103 (https://doi.org/10.5256/f1000research.175369.r345084)
NOTE: it is important to ensure the information in square brackets after the title is included in all citations of this article.
Views
6
Cite
Reviewer Report 04 Dec 2024
Konstantinos Kalimeris, University of Zurich, Zurich, Switzerland 
Approved
VIEWS 6
The authors have provided ... Continue reading
CITE
CITE
HOW TO CITE THIS REPORT
Kalimeris K. Reviewer Report For: Coiling of echogenic perineural catheters with integral stylet: A proof-of-concept randomized control trial in a sciatic nerve block simulator and a pilot study in orthopaedic-trauma patients. [version 2; peer review: 2 approved]. F1000Research 2024, 13:1103 (https://doi.org/10.5256/f1000research.175369.r345083)
NOTE: it is important to ensure the information in square brackets after the title is included in all citations of this article.
Version 1
VERSION 1
PUBLISHED 27 Sep 2024
Views
23
Cite
Reviewer Report 25 Nov 2024
Konstantinos Kalimeris, University of Zurich, Zurich, Switzerland 
Approved with Reservations
VIEWS 23
This paper presents an interesting method to coil perineural catheters which are not self-coiling. The authors conducted a study in Blue Phantom comparing if removing the stylet from the catheter before threading the catheter improves the coiling and then used ... Continue reading
CITE
CITE
HOW TO CITE THIS REPORT
Kalimeris K. Reviewer Report For: Coiling of echogenic perineural catheters with integral stylet: A proof-of-concept randomized control trial in a sciatic nerve block simulator and a pilot study in orthopaedic-trauma patients. [version 2; peer review: 2 approved]. F1000Research 2024, 13:1103 (https://doi.org/10.5256/f1000research.170554.r342696)
NOTE: it is important to ensure the information in square brackets after the title is included in all citations of this article.
  • Author Response 02 Dec 2024
    Eleni Poulogiannopoulou, Second Department of Anaesthesiology, National and Kapodistrian University of Athens School of Medicine, Athens, 15349, Greece
    02 Dec 2024
    Author Response
    We would like to thank you for the feedback and comments. In the revised manuscript, all recommendations were considered and incorporated in the new text.  Corrections in the text are highlighted ... Continue reading
COMMENTS ON THIS REPORT
  • Author Response 02 Dec 2024
    Eleni Poulogiannopoulou, Second Department of Anaesthesiology, National and Kapodistrian University of Athens School of Medicine, Athens, 15349, Greece
    02 Dec 2024
    Author Response
    We would like to thank you for the feedback and comments. In the revised manuscript, all recommendations were considered and incorporated in the new text.  Corrections in the text are highlighted ... Continue reading
Views
28
Cite
Reviewer Report 06 Nov 2024
Agathi Karakosta, University of Ioannina, Ioannina, Greece 
Approved
VIEWS 28
COMMENTS TO THE AUTHORS

First, I would like to congratulate the authors for the interesting and relevant work. Data is presented in a detailed and coherent manner.

However, there are minor issues that need ... Continue reading
CITE
CITE
HOW TO CITE THIS REPORT
Karakosta A. Reviewer Report For: Coiling of echogenic perineural catheters with integral stylet: A proof-of-concept randomized control trial in a sciatic nerve block simulator and a pilot study in orthopaedic-trauma patients. [version 2; peer review: 2 approved]. F1000Research 2024, 13:1103 (https://doi.org/10.5256/f1000research.170554.r327786)
NOTE: it is important to ensure the information in square brackets after the title is included in all citations of this article.
  • Author Response 02 Dec 2024
    Eleni Poulogiannopoulou, Second Department of Anaesthesiology, National and Kapodistrian University of Athens School of Medicine, Athens, 15349, Greece
    02 Dec 2024
    Author Response
    Thank you very much, indeed, for all your significant remarks. In the revised manuscript, all recommendations were considered and incorporated in the new text.  Corrections in the text are highlighted ... Continue reading
COMMENTS ON THIS REPORT
  • Author Response 02 Dec 2024
    Eleni Poulogiannopoulou, Second Department of Anaesthesiology, National and Kapodistrian University of Athens School of Medicine, Athens, 15349, Greece
    02 Dec 2024
    Author Response
    Thank you very much, indeed, for all your significant remarks. In the revised manuscript, all recommendations were considered and incorporated in the new text.  Corrections in the text are highlighted ... Continue reading

Comments on this article Comments (0)

Version 2
VERSION 2 PUBLISHED 27 Sep 2024
Comment
Alongside their report, reviewers assign a status to the article:
Approved - the paper is scientifically sound in its current form and only minor, if any, improvements are suggested
Approved with reservations - A number of small changes, sometimes more significant revisions are required to address specific details and improve the papers academic merit.
Not approved - fundamental flaws in the paper seriously undermine the findings and conclusions
Sign In
If you've forgotten your password, please enter your email address below and we'll send you instructions on how to reset your password.

The email address should be the one you originally registered with F1000.

Email address not valid, please try again

You registered with F1000 via Google, so we cannot reset your password.

To sign in, please click here.

If you still need help with your Google account password, please click here.

You registered with F1000 via Facebook, so we cannot reset your password.

To sign in, please click here.

If you still need help with your Facebook account password, please click here.

Code not correct, please try again
Email us for further assistance.
Server error, please try again.